  • Source One or Multiple RTD Curves
  • High Accuracy
  • Easy-to-Read Backlight Display
  • Calibrate Directly in Temperature (°C or °F)
  • Compatible with All RTD Process Instruments
Equipamentos de Precisão para Calibração e Referências de Ponto de Gelo - Ver Produtos Relacionados

Descrição

The CL510A Series can check and calibrate all your RTD instruments. Take it into the shop, plant or field without worry. It’s easy to operate even in the dark areas of the plant with the backlit display. Stop carrying around a decade box and RTD resistance tables, with the CL510A you can calibrate directly in temperature (°C and °F). Choose between a single RTD type model, CL510A-1 (Pt100, 0.00385), or a CL510A-10 for multiple RTD types.

The CL510A-10 works with a wide variety of RTD curves including Platinm 100 (alpha = 3850, 3902, 3916, 3926) and 1000 (alpha = 3850, 3750) Ω, Copper 10 and 50 Ω, Nickel 100 and 120 Ω. Easily set any value quickly to within 0.1° with the adjustable digital potentiometer "DIAL". Connect directly to RTD inputs of smart transmitters, PLCs, controllers and multichannel recorders and verify their displays.

SPECIFICATIONS (Unless otherwise indicated, all specifications are rated from a nominal 23°C (73°F), 70% RH)
Accuracy: ±(0.015% of Rdg + 0.05 Ω)
Temperature Drift: ±0.05 Ω/°C
Operating Temperature Range: -25 to 60°C (-10 to 140°F)
Relative Humidity Range, 10% ≤RH:
    ≤90%: 0 to 35°C (32 to 95°F), non-condensing
    ≤70%: 35 to 60°C (95 to 140°F), non-condensing
Size: 126 L x 69 W x 45 mm H (4.96 x 2.73 x 1.79")
Weight: 0.24 kg (8.4 oz) (including batteries)
Batteries: 4 "AA" alkaline 1.5V (included)
Battery Life: 50 hr
Low Battery: Low battery indication with nominal 1 hour of operation left
Protection Against Misconnection: Over-voltage protection to 60 Vdc (rated for 30 sec)
Display: High contrast graphic liquid crystal display, LED backlighting for use in low lit areas

Source
Accuracy: From 1 to 10.2 mA
External Excitation Current: ±(0.015% of rdg + 0.05 Ω) Below 1 mA of External
Excitation Current: ±(0.015% of rdg + 0.01 mV/mA excitation current + 0.05 Ω)
Resistance Ranges: 0.00 to 410.00, 410.1 to 4001.0 Ω
Allowable Excitation Current Range, Steady or Pulsed/Intermittent:
    <410 Ω: 10.2 mA max
    410 to 001 Ω: 1 mA max Pulsed Excitation Current
Compatibility: DC to 0.01 second pulse width