SPECIFICATIONS
Pressure:
Accuracy (1 year):
Includes all effects of linearity, hysteresis, repeatability and temperature
Port 1: ±(0.05% rdg +0.005% FS [36psi] Vacuum: ±(0.25% rdg +0.025% FS [-14.7 psi])
Scale/Resolution
0.01 inH2O
0.001 psi
0.01 kPa
1 mmH2O
0.01 mmHg
0.001 inHg
0.0001 kg/cm2
0.1 mbar
Port 2: ±(0.05% rdg + 0.005% FS [3000psi])
Scale/ Resolution
0.1 psi
0.01 Bar
1 kPa

Electrical:
Agency Approvals (Pending): Factory Mutual Intrinsically Safe IS/I/1/ABCD, IS/II/1/EG
Range: 0 to 55 mA
Resolution: 0.001 mA
Accuracy: ±(0.025% rdg +0.001 mA)
Percent Conversion: 4 mA = 0.00%, 20 mA = 100.00% 10 mA = 0.00%, 50 mA = 100.00%
RS-232 Comm: All Models

Temperature:
Operating:
0 to 50°C (32 to 122°F)
Compensated: 0 to 50°C (32 to 122°F)
Storage: -20 to 70°C (-4 to 158°F)

Physical:
Media Compatibility:
Liquids and gasses compatible with PTFE penetrated-hard anodized aluminum, 316 stainless steel and FKM
Power: 9 V alkaline included (low bat indication) standard 60 hr typical life without RS-232 40 hr typical life using RS-232
Pressure Connection: 1 /8 NPTF
RS232 Connection: 9-position PC card I/O receptacle. Mating Connector P/N 621370-1
Dimensions: 130 H x 70 W x 29 mm D (5.1 H x 2.8 W x 1.1" D)
Case: Extruded Aluminum
Weight: 385 g (13.5 oz)