High Accuracy Pressure Transmitter with Shunt Calibrator for Quick Calibration Checks

PX941A

High Accuracy Pressure Transmitter with Shunt Calibrator for Quick Calibration Checks | PX941A

imagem meramente ilustrativa

CE
  • Stainless Steel Diaphragm for Compatibility with Most Media
  • 1/4 NPT Fitting For Fast, Easy, and Secure Installation
  • Rugged Stainless Steel Case Protects Components from Industrial Environments
  • Quick Disconnect Style for Easy Field Connections
  • 2-wire, 4-20 mA Output Provides Excellent Noise Immunity over Long Cable Runs
  • Adjustable Zero and Span
  • Units <100 psi are Vented Gage for High Precision
Transdutores de Pressão - Ver Produtos Relacionados

Descrição



SPECIFICATIONS
Excitation: 12 to 40 Vdc @ 20 mA
Output: 4 to 20 mA (2 wire) (±10% adjustable)
Insulation Resistance: 5 MΩ @ 75 Vdc PERFORMANCE
Accuracy: ±0.15% (combined effect of linearity, hysteresis and repeatability)
Shunt Value: On calibration sheet (80% FS)
Hysteresis: ±0.10% typical
Repeatability: ±0.05% typical
Zero Balance: 4 mA (+10%, -2% adj)
Operating Temperature Range: -18 to 85°C (0 to 185°F)
Compensated Temperature Range: 16 to 71°C (60 to 160°F)
Thermal Zero Effect: ±0.005% full scale/°F
Thermal Sensitivity Effect: ±0.005% reading/°F
Proof Pressure: 150% range
Burst Pressure: 300% range CONSTRUCTION
Body Material: 17-4 PH stainless steel
Diaphragm Material: 17-4 PH stainless steel
Pressure Port: 1/4-18 NPT
Male Thread: 10 to 1000 psi
Female thread: 3000 to 7500 psi
Mating Connector: PT06F10-6S (not included)
Weight: 383 g (13.5 oz)