Super OMEGACLAD™ XL Thermocouple Probes with Mini Transition - Super Accurate, Super Stable

KMTXL and NMTXL

Super OMEGACLAD™ XL Thermocouple Probes - Miniature Transition Junction Probe | KMTXL and NMTXL

imagem meramente ilustrativa

RoHS
  • Very Low Drift at High Temperatures
  • Super Accurate, Super Stable, Super Value
  • For Use Up to 1335°C (2440°F) (2435°F) Short Duration Capability 1.5 mm and Larger Diameters
  • 1150°C (2100°F) Continuous Use
    3.0 mm and Larger Diameters 4,500 hours (25 plus weeks) *Average MTBF (Mean Time Between Failure)
    1.5 mm (0.062") 2,000 Hours (12 weeks)* Average MTBF (Mean Time Between Failure)
    *Performance Data Established by Means of an Electric Furnace at Static 1150°C (2100°F) Temperature in an Environment. The Data is for Reference Purposes only and Does not Imply a Period of Guarantee
  • Excellent Long Term Stability at High Temperatures
  • Made From Special Limits of Error Material
  • Nickel-Chrome Based Super OMEGACLAD™ Sheathing Provides Excellent High Temperature Performance
  • Excellent Oxidation, Carburization and Chlorination Resistance
  • Probes include 36 inches of lead wire
Termopares - Ver Produtos Relacionados

Descrição



New Super OMEGACLAD™ XL sheathed mineral-insulated thermocouple wire, designated “XL” for ordering, can be used at ultra-high temperatures for prolonged periods with very little drift of the base metal thermocouple. The sheath material uses Nickel-Chrome based Super OMEGACLAD™ XL sheathing, which provides excellent oxidation resistance. The sheath can withstand exposure to combustion gases or air at temperatures up to 1335°C (2435°F). Superior resistance to oxidation attack results from a tenacious and protective high-temperature film that does not affect the stability of the thermocouple alloys. This film permits the sensor to be used at ultra-high temperatures for prolonged periods with improved accuracy over other available sheath materials. Super OMEGACLAD™ XL sheathing also provides excellent resistance to corrosion in high temperature chlorine-contaminated oxidizing environments and ammonia/nitride-rich environments at temperatures above 980°C (1800°F), the temperature at which the protective film forms.

See the Related Links section at the bottom of this page for addtional models.




Standard Dimension - Miniature Transition Junction Probes
Alloy/ANSI
Color Code
Sheath Dia.
inches
Model Number
ANSI Color Code
K
CHROMEGA™-ALOMEGA™
XL Sheath
0.010"
0.020"
0.032"
0.040"
0.062"
0.125"
KMTXL-010[*]-[**]
KMTXL-020[*]-[**]
KMTXL-032[*]-[**]
KMTXL-040[*]-[**]
KMTXL-062[*]-[**]
KMTXL-125[*]-[**]
N
OMEGA-P™-OMEGA-N™
XL Sheath
0.020"
0.032"
0.040"
0.062"
0.125"
NMTXL-020[*]-[**]
NMTXL-032[*]-[**]
NMTXL-040[*]-[**]
NMTXL-062[*]-[**]
NMTXL-125[*]-[**]
[*]Specify junction type: G (Grounded), E (Exposed), or U (Ungrounded).
[**]Specify Length in inches: 6, 12 ,18, or 24



Click here for help in selecting the thermocouple junction.


View the Thermocouple Accuracy and Color Code charts
We noticed you are from the US. Visit your local site for regional offers and live support.